《Otology and neurotology: official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology》 2022年43卷7期
摘要
:
Objective Monitoring four-point impedance changes after cochlear implantation with comparison to conventional impedance measurements. Four-point impedance provides information regarding the bulk biological environment surrounding ...
展开
Objective Monitoring four-point impedance changes after cochlear implantation with comparison to conventional impedance measurements. Four-point impedance provides information regarding the bulk biological environment surrounding the electrode array, which is not discernible with conventional impedances. Study Design Prospective observational. Setting Hospital. Patients Adult cochlear implant recipients with no measurable hearing before implantation and implanted with a perimodiolar cochlear implant. Main Outcome Measure(s) Mean values for four-point and common ground impedances were calculated for all electrode contacts at intra-operative, 1 day, 1 week, 4 to 6 weeks, and 3 months post implantation. Linear mixed models were applied to the impedance data to compare between impedances and time points. Furthermore, patients were divided into groups dependent on the normalized change in four-point impedance from intra-operative to 1 day post-operative. The normalized change was then calculated for all other time points and compared across the two groups. Results Significant increases in four-point impedance occurred 1 day and 3 months after surgery, particularly in the basal half of the array. Four-point impedance at 1 day was highly predictive of four-point impedance at 3 months. Four-point impedance at the other time points showed marginal or no increases from intra-operative. Patients with an average increase higher than 10% in four-point impedance from intra-operative to 1 day, had significantly higher values at 3 months (p = 0.012). These patterns were not observed in common ground impedance. Conclusion This is the first study to report increases in four-point impedance within 24 hours of cochlear implantation. The increases at 1 day and 3 months align with the natural timeline of an acute and chronic inflammatory responses.
收起